JP | JPY
-
-
-
- Challenges and Chances: A Review of the 1st Stem Cell Community Day
- Summertime, and the Livin’ Is Easy…
- Follow-on-Biologics – More than Simple Generics
- Bacteria Versus Body Cells: A 1:1 Tie
- Behind the Crime Scene: How Biological Traces Can Help to Convict Offenders
- Every 3 Seconds Someone in the World Is Affected by Alzheimer's
- HIV – It’s Still Not Under Control…
- How Many Will Be Convicted This Time?
- Malaria – the Battle is Not Lost
- Physicians on Standby: The Annual Flu Season Can Be Serious
- At the Forefront in Fighting Cancer
- Molecular Motors: Think Small and yet Smaller Again…
- Liquid Biopsy: Novel Methods May Ease Cancer Detection and Therapy
- They Are Invisible, Sneaky and Disgusting – But Today It’s Their Special Day!
- How Many Cells Are in Your Body? Probably More Than You Think!
- What You Need to Know about Antibiotic Resistance – Findings, Facts and Good Intentions
- Why Do Old Men Have Big Ears?
- The Condemned Live Longer: A Potential Paradigm Shift in Genetics
- From Research to Commerce
- Chronobiology – How the Cold Seasons Influence Our Biorhythms
- Taskforce Microbots: Targeted Treatment from Inside the Body
- Eyes on Cancer Therapy
-
-
-
-
-
- Challenges and Chances: A Review of the 1st Stem Cell Community Day
- Summertime, and the Livin’ Is Easy…
- Follow-on-Biologics – More than Simple Generics
- Bacteria Versus Body Cells: A 1:1 Tie
- Behind the Crime Scene: How Biological Traces Can Help to Convict Offenders
- Every 3 Seconds Someone in the World Is Affected by Alzheimer's
- HIV – It’s Still Not Under Control…
- How Many Will Be Convicted This Time?
- Malaria – the Battle is Not Lost
- Physicians on Standby: The Annual Flu Season Can Be Serious
- At the Forefront in Fighting Cancer
- Molecular Motors: Think Small and yet Smaller Again…
- Liquid Biopsy: Novel Methods May Ease Cancer Detection and Therapy
- They Are Invisible, Sneaky and Disgusting – But Today It’s Their Special Day!
- How Many Cells Are in Your Body? Probably More Than You Think!
- What You Need to Know about Antibiotic Resistance – Findings, Facts and Good Intentions
- Why Do Old Men Have Big Ears?
- The Condemned Live Longer: A Potential Paradigm Shift in Genetics
- From Research to Commerce
- Chronobiology – How the Cold Seasons Influence Our Biorhythms
- Taskforce Microbots: Targeted Treatment from Inside the Body
- Eyes on Cancer Therapy
-
-
JP | JPY
epMotion®分注ツール
製品情報
epMotionリキッドハンドリングワークステーションの分注ツールは、さまざまな液体をピペッティングしてディスペンスするための精密機器です。それらはシングルチャンネルツールとしてもマルチチャンネルツールとしても利用可能であり、それらは0.2~1,000 µLの範囲の容量をカバーしています。追加でダウンロードできる材料がありますこのページの最後に
製品 (12)
12 製品
|
|
|
|
|
|
|
|
|
|
製品をもっと見せる
お客様のショッピングカートに追加されました。
製品情報
テクニカルデータ
特徴
製品情報
epMotionリキッドハンドリングワークステーションの分注ツールは、さまざまな液体をピペッティングしてディスペンスするための精密機器です。 それらはシングルチャンネルツールとしてもマルチチャンネルツールとしても利用可能であり、それらは0.2~1,000 µLの範囲の容量をカバーしています。
テクニカルデータ
仕様 | TS 10 | TS 50 | TS 300 | TS 1000 | TM 10-8 | TM 50-8 | TM 300-8 | TM 1000-8 |
---|
チャネル数 | 1-チャネル | 1-チャネル | 1-チャネル | 1-チャネル | 8-チャネル | 8-チャネル | 8-チャネル | 8-チャネル |
容量レベル | 0.2 – 10 µL | 1.0 – 50 µL | 20 – 300 µL | 40 – 1,000 µL | 0.2 – 10 µL | 1.0 – 50 µL | 20 – 300 µL | 40 – 1,000 µL |
テスト容量 |
0.2 µL 1 µL 5 µL 10 µL |
1 µL 5 µL 25 µL 50 µL |
20 µL 30 µL 150 µL 300 µL |
40 µL 100 µL 500 µL 1000 µL |
1 µL 5 µL 10 µL |
1 µL 5 µL 25 µL 50 µL |
20 µL 30 µL 150 µL 300 µL |
40 µL 100 µL 500 µL 1000 µL |
相対系統誤差 |
±25 % ±5 % ±2.4 % ±1.2 % |
±15 % ±5.0 % ±1.5 % ±1.0 % |
±4.0 % ±3.0 % ±1.0 % ±0.6 % |
±5.0 % ±2.0 % ±1.0 % ±0.7 % |
±7.5 % ±2.5 % ±2.0 % |
±25.0 % ±5.0 % ±2.0 % ±1.2 % |
±10.0 % ±10.0 % ±2.5 % ±1.5 % |
±6.0 % ±3.0 % ±1.5 % ±0.8 % |
絶対系統誤差 |
±0.05 µL ±0.05 µL ±0.12 µL ±0.12 µL |
±0.15 µL ±0.25 µL ±0.375 µL ±0.5 µL |
±0.8 µL ±0.9 µL ±1.5 µL ±1.8 µL |
±2.0 µL ±2.0 µL ±5.0 µL ±7.0 µL |
±0.075 µL ±0.125 µL ±0.02 µL |
±0.25 µL ±0.25 µL ±0.5 µL ±0.6 µL |
±2.0 µL ±3.0 µL ±3.75 µL ±4.5 µL |
±2.4 µL ±3.0 µL ±7.5 µL ±8.0 µL |
絶対ランダム誤差 |
±0.04 µL ±0.03 µL ±0.025 µL ±0.025 µL |
±0.05 µL ±0.15 µL ±0.15 µL ±0.2 µL |
±0.5 µL ±0.45 µL ±0.6 µL ±0.9 µL |
±0.6 µL ±1.0 µL ±1.0 µL ±1.5 µL |
±0.05 µL ±0.1 µL ±0.06 µL |
±0.1 µL ±0.25 µL ±0.3 µL ±0.3 µL |
±0.8 µL ±1.05 µL ±1.2 µL ±1.5 µL |
±1.0 µL ±1.5 µL ±1.5 µL ±1.5 µL |
相対ランダム誤差 |
±19.8 % ±3 % ±0.5 % ±0.25 % |
±5.0 % ±3.0 % ±0.6 % ±0.4 % |
±2.5 % ±1.5 % ±0.4 % ±0.3 % |
±1.5 % ±1.0 % ±0.2 % ±0.15 % |
±5 % ±2 % ±0.6 % |
±10.0 % ±5.0 % ±1.2 % ±0.6 % |
±4.0 % ±3.5 % ±0.8 % ±0.5 % |
±2.5 % ±1.5 % ±0.3 % ±0.15 % |
特徴
- 200 nL~1,000 µLの最高精度
- 各種液体用
- 1チャンネルと8チャンネル分注ツールレンタル
ダウンロード: epMotion®分注ツール
Does a service carried out by Eppendorf meet the GMP requirement?
Eppendorf offers global certification services which comply with FDA 21 CFR Part 11, EU GMP Annex 11, GLP and GAMP®5 to ensure a consistent, standardized operation. We also enable an uncomplicated validation from the implementation to the shutdown by providing support for the installation und operational qualifications (IQ/OQ) of the device and with the software. The entire qualification is documented in detail and the operator’s personnel is qualified and certified in intensive training sessions.
もっと読む
読む量をもっと少なくする
How much are the different epMotion service programs?
The costs depend on the model and service that is performed. Please contact your local Eppendorf Service.
もっと読む
読む量をもっと少なくする
How often do dispensing tools need to be calibrated?
As part of his test equipment monitoring process, the operator determines the frequency of the calibration as well as the quantity of the testing volume and the number of measurements for each testing volume. This may vary depending on how often the device is used, by how many operators it is used, the aggressivity of the liquids used and the maximum permissible errors defined by the operator. The dispensing tools of your device will display after 200,000 runs that an inspection is recommended. However, there should be at least 1 calibration per year.
もっと読む
読む量をもっと少なくする
How often does my epMotion need to be serviced?
The need for service depends on how intensively the device is used and if you are working in a regulated field with special service demands. However, we recommend to have it serviced at least once a year.
もっと読む
読む量をもっと少なくする
How often does the OQ need to be carried out for an epMotion?
The company is responsible for technical checks of devices. Eppendorf recommends regular device qualifications depending on the intensity of use, however, they should be carried out at least once a year.
もっと読む
読む量をもっと少なくする
What do you need to look out for before carrying out a service?
The device needs to be decontaminated first. The device must be freely accessible and all samples removed from the device.
もっと読む
読む量をもっと少なくする
What is IQ or OQ and what do I need them for?
Installation and operational qualification; the IQ documents that the device has been delivered and installed correctly. The OQ checks the correct functioning of a device. IQ/OQ ensure that your quality management requirements are met and that your device operates properly and in accordance with the manufacturer’s specifications.
もっと読む
読む量をもっと少なくする
What types of service programs are available for epMotion?
各器機に提供されているサービスプログラムは、サービス&サポート/メンテナンスおよび証明書サービスセクションにあります。
もっと読む
読む量をもっと少なくする
What types of training or online training do you offer to enable me to improve my skills?
Please visit the Eppendorf webpage about the Eppendorf Training Center (https://www.eppendorf.com/training or contact your local Eppendorf Service for further information and a quotation.
もっと読む
読む量をもっと少なくする
Which consumables should be replaced at regular intervals?
All three XYZ flex cables should be replaced at least once per year if the machine is used daily.
もっと読む
読む量をもっと少なくする
Who is my local service contact?
Please contact your local Eppendorf Service Team. You find the contact data on our homepage (eppendorf.com/contact) or in the information materials.
もっと読む
読む量をもっと少なくする
Why can the dispending tool be calibrated without the user unit ?
The epMotion dispensing tool is calibrated with the help of the epMotion Tool Holder. The electronic parameter to control the tools of the epMotion are always the same in all epMotion units. Therefore all epMotion dispensing tool can be calibrated on different units than the user unit. The only exceptions are the TS10 and TM10 dispensing tool. They can only be calibrated with units controlled by epBlue Version 40.5.
もっと読む
読む量をもっと少なくする
Why do the dispensing tools of my epMotion need to be calibrated?
The dispensing tools are precision devices. Servicing and calibration at regular intervals ensure a consistently high precision of the results. We provide documentation in support for GLP to assist you.
もっと読む
読む量をもっと少なくする
Why should I have my epMotion serviced at regular intervals?
To ensure the optimum performance and precision of your device. This also enables the early detection of any defects which can then be corrected easily. Only precise and accurate devices ensure successful liquid transfers!
もっと読む
読む量をもっと少なくする