text.skipToContent text.skipToNavigation
No results found
Search Suggestions

DASGIP® Parallel Bioreactor System

Catalog No. 76DG04CCBB

Product Information

  • for cell culture, max. 50 sL/h gassing
  • 4-fold system with DASGIP® Bioblock

You will find additional download material at the bottom of this page

Technical Data

Features

Specifications DASGIP® System for Cell Culture
Application Cell culture 
Number of parallel units Up to 16 
Software DASware control, other DASware optional 
User Interface Process computer with monitor 
Typical power consumption (4-fold system without process computer) 480 W (at 230 V)/
432 W (at 115 V)
(595 W/552 W without DASGIP Bioblock) 
Dimensions (W × D × H, 4-fold system without process computer) 140 x 70 x 53 cm (55 x 28 x 21 in) 
Typical weight 78 kg
(60 kg without DASGIP Bioblock) 
Bacteria/Yeast/Fungi yes 
Mammalian/Animal yes 
Insect yes 
Interchangeable Vessels yes 
Validatable no 
Stem cells yes 

Bioreactors

Vessels Glass/single-use vessels 
Sterilization Autoclavable/
pre-sterilized 
Total volume 1.5 L to 2.3 L 

Agitation

Drive Direct/magnetic overhead drive 
Speed ranges 30 – 1,250 rpm (standard)/
100 – 1,600 rpm (optional) 
Impellers Pitched-blade 

Gassing

Gas supply TMFC; overlay and/or sparger 
Standard gas flow rates 0.1 – 50 sL/h, 0.1 – 40 sL/h CO2 (DASGIP MX4/4, other modules available) 
Standard gas mixing Air, N2, O2 and/or CO2 

Feeding

Feed lines per vessel Up to 8 
Standard feed rates (depending on tube diameter) 0.3 – 9.5 mL/h to 13 – 420 mL/h (DASGIP MP8) 10 – 70 mL/h to 0.4 – 5 L/h
(DASGIP MP4) 

Monitoring and Control

Temperature control Heating and cooling integrated in DASGIP Bioblock (additional cooling fingers optional) 
Standard temperature range (at 25° C room temp.) 5 K above cooling agent temperature – 99°C (DASGIP Bioblock)1) 
pH control
DO control Cascade (O2 concentration, gas flow rate), and other set-ups 
ORP (redox) measurement
Level/foam Optional 
OD measurement Optional
(DASGIP OD4) 
Exhaust condensation Water-cooled or liquid-free (Peltier with DASGIP EGC4) 
Exhaust analysis Optional
(DASGIP GA4) 

1)Application-specific operating conditions apply.