text.skipToContent text.skipToNavigation
No results found
Search Suggestions

Accessories (4)

Spare Parts (2)


    • 1-channel
    • 1 µL – 10 mL
    • incl. 1x Combitips® advanced 2.5 mL (green) tip and Pipette Holder 2
    • ACT-labeled: Independent third-party product sustainability validation by My Green Lab
    More Product Details ...
    Catalog no. 4982000012

      • contains one Combitips® advanced sample of each size and Adapter advanced 25 and 50 mL
      • Eppendorf Quality
      More Product Details ...
      Catalog no. 0030089936

        • 1-channel
        • 1 µL – 50 mL
        • incl. charging cable and Combitips® advanced assortment pack (1 tip of each size)
        • ACT-labeled: Independent third-party product sustainability validation by My Green Lab
        More Product Details ...
        Catalog no. 4987000029

          • 1-channel
          • 1 µL – 50 mL
          • incl. charging cable and Combitips® advanced assortment pack (1 tip of each size)
          • ACT-labeled: Independent third-party product sustainability validation by My Green Lab
          More Product Details ...
          Catalog no. 4987000010

          Technical Data

          Features

          Specifications Combitips® advanced - 0030089839
          ACT label yes 
          Product type Consumables 
          Pipetting type Positive displacement system 
          Tip size 25 mL 
          Color code red 
          Tip color colorless tips 
          Purity grade PCR clean 
          Packaging 100 pcs. (4 reclosable bags × 25 pcs.), each box containing one adapter 
          Open system (third party consumables) no 

          Performance data with Multipette E3 & E3x / Repeater E3 & E3x

          Random measurement error 250 µL: ± 3%; ± 7.5 µL
          2500 µL: ± 0.35%; ± 8.75 µL
          12500 µL: ± 0.25%; ± 31.25 µL
          25000 µL: ± 0.15%; ± 37.5 µL 
          Systematic measurement error 250 µL: ± 2.5%; ± 6.25 µL
          2500 µL: ± 0.3%; ± 7.5 µL
          12500 µL: ± 0.3%; ± 37.5 µL
          25000 µL: ± 0.3%; ± 75 µL 
          Volume range 250 – 25,000 µL  

          Performance data with Multipette M4 / Repeater M4

          Random measurement error 250 µL: ± 3%; ± 7.5 µL
          500 µL: ± 0.6%; ± 3 µL
          2500 µL: ± 0.5%; ± 12.5 µL
          5000 µL: ± 0.25%; ± 12.5 µL 
          Systematic measurement error 250 µL: ± 3%; ± 7.5 µL
          500 µL: ± 0.4%; ± 2 µL
          2500 µL: ± 0.3%; ± 7.5 µL
          5000 µL: ± 0.3%; ± 15 µL 
          Volume range 250 – 5,000 µL