- Combitips® advanced tips are the dedicated consumable for all Multipette® multi-dispenser pipettes from Eppendorf
- The 9 volume sizes (0.1 mL – 50 mL) offer a maximum range of dispensing volumes with increments as low as 100 nanoliters depending on the Combitips® advanced used
- Combitips® advanced in ‘Biopur®’ quality are guaranteed to be sterile and free of pyrogens, RNases, DNases, DNA and ATP as well as PCR inhibitors
- Lot-specific certificates for Eppendorf consumables in ‘Biopur®’ quality are available at www.eppendorf.com/certificates
- Elongated Combitips® advanced in volumes 2.5 mL, 5.0 mL and 10 mL make it possible to reach the bottom of most common laboratory vessels
- An individual color coding on each tip facilitates the quick identification of the desired Combitips® advanced dispenser tip
- The unique funnel geometry ensures comfortable handling while preventing damage to gloves
- ACT®-labeled by non-profit My Green Lab® to help you make an informed product choice when it comes to sustainability
Combitips® advanced
Product Information
- Biopur®
- 10 mL, orange, colorless tips
- 100 pcs., individually blister-wrapped
You will find additional download material at the bottom of this page
Accessories (5)
|
|
|
|
|
Technical Data
Features
Technical Data
Specifications | Combitips® advanced - 0030089677 |
---|
ACT label | yes |
Product type | Consumables |
Pipetting type | Positive displacement system |
Tip size | 10 mL |
Color code | orange |
Tip color | colorless tips |
Purity grade | Biopur® |
Packaging | 100 pcs., individually blister-wrapped |
Open system (third party consumables) | no |
Performance data with Multipette E3 & E3x / Repeater E3 & E3x |
|
Random measurement error |
100 µL: ± 3.5%; ± 3.5 µL 1000 µL: ± 0.25%; ± 2.5 µL 5000 µL: ± 0.25%; ± 12.5 µL 10000 µL: ± 0.15%; ± 15 µL |
Systematic measurement error |
100 µL: ± 1.5%; ± 1.5 µL 1000 µL: ± 0.5%; ± 5 µL 5000 µL: ± 0.4%; ± 20 µL 10000 µL: ± 0.4%; ± 40 µL |
Volume range | 100 – 10.000 µL |
Performance data with Multipette M4 / Repeater M4 |
|
Random measurement error |
100 µL: ± 4%; ± 4 µL 200 µL: ± 0.6%; ± 1.2 µL 1000 µL: ± 0.4%; ± 4 µL 2000 µL: ± 0.25%; ± 5 µL |
Systematic measurement error |
100 µL: ± 3%; ± 3 µL 200 µL: ± 0.5%; ± 1 µL 1000 µL: ± 0.5%; ± 5 µL 2000 µL: ± 0.5%; ± 10 µL |
Volume range | 100 – 2.000 µL |
Downloads: Combitips® advanced