- Mechanical air-cushion pipette with outstanding precision and accuracy - operated with just one button for volume selection, pipetting and tip ejection
- With its innovative one-button operation, the Reference 2 allows fast and ergonomic handling with reduced operating efforts and active aerosol reduction. A clear haptic feedback helps to avoid accidental tip ejection
- 4-digit display with magnifying window for quick and secure volume setting and identification (clearly visible from every angle), incl. volume lock
- Spring-loaded tip cones improve user-to-user reproducibility and ergonomics by lowering the tip attachment force
- The outstanding precision and accuracy of Reference 2 provides reliable results
- A temporary adjustment allows for an easy adjustment of your Reference 2 for the most accurate pipetting of different liquids or in changing external conditions without the need for a full calibration
- The Reference 2 multi-channel variants offer high flexibility incl. a spring-loaded tip cone which can be switched on/off optionally
No results found
Search Suggestions
Eppendorf Reference® 2
Product Information
- 8-channel
- 10 – 100 µL, yellow, variable volume
- incl. epT.I.P.S.® Box 2.0 with 96 pipette tips
You will find additional download material at the bottom of this page
Pipette Tips (32)
Services (11)
Accessories (6)
Spare Parts (7)
Catalog no.
0030078543
|
Catalog no.
0030078713
|
Catalog no.
0030000889
|
Catalog no.
0030073436
|
Catalog no.
0030073819
|
Catalog no.
0030076150
|
Catalog no.
0030076320
|
Catalog no.
0030010043
|
Catalog no.
0030075234
|
Catalog no.
0030000870
|
Added to Your Cart
Technical Data
Features
Technical Data
Specifications | Eppendorf Reference® 2 - 4926000034 |
---|
Product type | Pipettes |
Pipetting type | Air-cushion system |
Number of channels | 8-channel |
Volume range | 10 – 100 µL |
Color code | yellow |
Volume increment | 0.1 µL |
Operating mode | mechanical |
Spring-loaded tip cone | yes |
Random measurement error |
10 µL: ± 2%; ± 0.2 µL 50 µL: ± 0.8%; ± 0.4 µL 100 µL: ± 0.3%; ± 0.3 µL |
Systematic measurement error |
10 µL: ± 3%; ± 0.3 µL 50 µL: ± 1%; ± 0.5 µL 100 µL: ± 0.8%; ± 0.8 µL |
Autoclavable | yes |
Consumables | epT.I.P.S.® |
Open system (third party consumables) | yes |
Delivery content | Pipette, Operating manual, Certificate, Adjustment seal for temporary adjustment, Adjustment seal for factory adjustment (x2), Grease for pipettes, O-ring tool |