text.skipToContent text.skipToNavigation
No results found
Search Suggestions

Centrifuge CR30NX

Catalog No. 5721300012

Product Information

  • refrigerated
  • 230 V/50/60 Hz (no plug), 30 A, 1-phase

You will find additional download material at the bottom of this page

Technical Data

Features

Specifications CR30NX
Max. RCF 110000 × g
Max. RCF with fixed-angle rotor 110000 × g
Max. RCF with swing-bucket rotor 110000 × g
Heat output 1.2 kW
Speed 300 – 30000 rpm 
Max. capacity 6 × 1,000 mL/ 4 × 1,500 mL 
Number of programs 1,000 programs 
Timer 1 s to 99 min 59 s (in increments of 1 s) with HOLD function or 1 min to 99 h 59 min (in increments of 1 min) with HOLD function 
Noise level < 56 dBA 
Interfaces 1xUSB, 1xLAN 
Power supply 230 V, 50/60 Hz 
Max. power consumption 1500 W
Dimensions (W × D × H) 708 × 892 × 866 mm 
Footprint (dimensions w/o front panel, WxD) 708 x 892 mm 
Product weight 310 kg 
Cooling refrigerated 
Separation example Very high centrifugation speed of up to 110,000 x g with fixed-angle rotor R30AT (30,000 rpm) and swing-bucket rotor R25ST (25,000 rpm) 
Temperature accuracy ±2 °C 
Acceleration/Deceleration mode Acceleration: 11 modes
Deceleration: 10 modes with 2 modes variable deceleration slope 
Drive unit Induction Motor (Inverter) 
Refrigerator Rotary Inverter Driven Compressor (Refrigerant: R410A) 
Keypad Touch sensitive LCD panel with GUI icons 
Safety devices Door interlock, overspeed detection, imbalance detection, rotor cover detection, over-temperature detection 
Power requirement Single phase: AC200/208/220/230/240 V ±10%, 50/60 Hz, 30 A
Three-phase four-wire: AC 380/400/415 V , ±10%, 50/60 Hz, 2 x 16 A1) 
User lockout function yes 
Operating Log Management yes 
Economy mode yes 
Continuous flow rotor yes 
Vacuum system yes 
Rotor stop signal select from 5 melodies, beep and mute 

1)Unbalanced three-phase (two-phase used). Neutral connection required.