text.skipToContent text.skipToNavigation
No results found
Search Suggestions

Dispensing head 96-channel 0.5 – 300 µL

  • New

Catalog No. 5069000527

Product Information

  • 96-channel, for epMotion® 96 Flex in hard shell case, needs base device for operation
  • 0.5 – 300 µL

You will find additional download material at the bottom of this page

Pipette Tips (6)


    • without filter
    • Eppendorf Quality, 0.5 – 50 µL, ACT, 2,304 tips (24 trays × 96 tips)
    More Product Details ...
    Catalog no. 0030014421

      • with filter
      • PCR clean, 0.5 – 50 µL, ACT, 2,304 tips (24 trays × 96 tips)
      More Product Details ...
      Catalog no. 0030014430

        • without filter
        • Eppendorf Quality, 5 – 300 µL, ACT, 2,304 tips (24 trays × 96 tips)
        More Product Details ...
        Catalog no. 0030014464

          • with filter
          • PCR clean and sterile, 0.5 – 50 µL, ACT, 2,304 tips (24 trays × 96 tips)
          More Product Details ...
          Catalog no. 0030014529

            • with filter
            • PCR clean, 5 – 300 µL, ACT, 2,304 tips (24 trays × 96 tips)
            More Product Details ...
            Catalog no. 0030014472

              • with filter
              • PCR clean and sterile, 5 – 300 µL, ACT, 2,304 tips (24 trays × 96 tips)
              More Product Details ...
              Catalog no. 0030014537

              Technical Data

              Features

              Specifications Dispensing head 0.5-300µL - 5069000527
              Pipetting type Air-cushion system 
              Number of channels 96-channel 
              Application Cell culture, Molecular biology 
              Volume range 0.5 – 300 µL 
              Operating mode
              Autoclavable no 
              Power supply
              Consumables epT.I.P.S.® Motion Reload System 
              Max. power consumption
              Dimensions (W × D × H)
              Product weight 2.1 kg / 4.6 lb
               
              Testing volume 50 µL Tips
              5 µL
              25 µL
              50 µL
              300 µL Tips
              30 µL
              150 µL
              300 µL 
              Systematic measurement error (determined in pipette mode) 50 µL Tips
              ≤ 3 % (not guaranteed, individual adjustment of device and dispensing head may be required)
              ≤ 4.5 %
              ≤ 1.8 %
              ≤ 1.2 %
              300 µL Tips
              ≤ 3 % (not guaranteed, individual adjustment of device and dispensing head may be required)
              ≤ 2.5 %
              ≤ 1.5 %
              ≤ 1.5 % 
              Random measurement error (determined in pipette mode) 50 µL Tips
              ≤ 2.5 % at 5 µL
              300 µL Tips
              ≤ 1.5 % at 30 µL 
              SLAS/ANSI positions

              Downloads: Dispensing head 96-channel 0.5 – 300 µL