- Decontamination from the outside
- Calibration test of 3 volumes, 4 measurements per volume in the 'as left condition'
- Preventive maintenance – internal & external cleaning, leak testing, lubricating of internal parts if applicable, and repair (parts replacements at an additional cost)
- Issuing of service report, test report and calibration label in compliance with ISO 8655 series regulations
- In the event of an adjustment or repair, a 3x10 measurement must be carried out in accordance with the ISO 8655:2022 standard. Therefore, we will perform and invoice a 3x10 measurement after each adjustment or repair, even if you have selected a calibration service with less measurements
- Please note: the next service due date will default to 12 months cycle unless otherwise specified/requested.
ID | IDR
ID | IDR
3x4 Calibration including Preventive Maintenance – 24 channels
Product Information
24-channel, pipette maintenance and calibration serviceYou will find additional download material at the bottom of this page
Features