text.skipToContent text.skipToNavigation
No results found
Search Suggestions

    3x4 Calibration including Preventive Maintenance – 8 channels

    Catalog No. 0082000163

    Product Information

    8-channel, pipette maintenance and calibration service

    You will find additional download material at the bottom of this page

    Features

    • Decontamination from the outside
    • Calibration test of 3 volumes, 4 measurements per volume in the 'as left condition'
    • Preventive maintenance – internal & external cleaning, leak testing, lubricating of internal parts if applicable, and repair (parts replacements at an additional cost)
    • Issuing of service report, test report and calibration label in compliance with ISO 8655 series regulations
    • In the event of an adjustment or repair, a 3x10 measurement must be carried out in accordance with the ISO 8655:2022 standard. Therefore, we will perform and invoice a 3x10 measurement after each adjustment or repair, even if you have selected a calibration service with less measurements
    • Please note: the next service due date will default to 12 months cycle unless otherwise specified/requested.
    Pipettes and dispensers and package equipment need to be decontaminated before sending in.

    Downloads: 3x4 Calibration including Preventive Maintenance – 8 channels

    Content