- Autoclavable and calibrated based on ISO 8655-6
- Compatible only with epMotion® software epBlue™ 40.6.2.6 (Firmware 15.09) or higher
ID | IDR
ID | IDR
TM 10-8 eight-channel dispensing tool
Technical Data
Features
Technical Data
Specifications | TM 10-8 |
---|
Number of channels | 8-channel |
Volume range | 0,2 – 10 µL |
Testing Volume |
1 µL 5 µL 10 µL |
Rel. systematic error |
±7.5 % ±2.5 % ±2.0 % |
Abs. systematic error |
±0.075 µL ±0.125 µL ±0.02 µL |
Abs. random error |
±0.05 µL ±0.1 µL ±0.06 µL |
Rel. random error |
±5 % ±2 % ±0.6 % |