- Decontamination from the outside
- Cleaning and lubrication of dispensing head
- Checking and exchanging of 96 O-rings on tip cones
- Leak test with 50 µl + 300 µl, or 300 µl + 1000 µl tips
- Gravimetric calibration test of 6 volumes, 5 measurements per volume in 'as left' condition
- Photometric test with 2 volumes, 5 measurements per volume in 'as left' condition
- Issuing of calibration test report and calibration label in compliance with ISO 23783 regulations
- Please note: the next service due date will default to 12 months cycle unless otherwise specified/requested
HK | HKD
HK | HKD
epMotion® 96 Flex dispensing head calibration, 5 measurements per test volume (min, mid, max each with smaller and larger tip) with preventive maintenance
Product Information
in accordance with ISO 23783, as-left calibration - 6x5You will find additional download material at the bottom of this page