- Bottle-top dispenser with standard GL 45 thread for most common bottle threads in all sizes
- Safety valve balls in discharge valve prevent leakage when discharge tube is not mounted
- Air vent cap can be unscrewed for easy assembly of drying tube
- Oval-shaped dispenser housing for easy transportation
- PFA seal of the sliding piston prevents seizing up
- Fixation slide on interior toothed track for rapid and safe volume adjustment
- High chemical resistance of all dispensing components
- Telescopic aspirating tubes for variable adjustment to bottle height
- Fully autoclavable without disassembly
- User adjustment option
HK | HKD
HK | HKD
Varispenser® 2x
Eppendorf Varispenser® 2 and Varispenser® 2x bottle-top dispensers – six different sizes for dispensing volumes from 0.2 – 100 mL
Varispenser® 2 bottle-top dispensers from Eppendorf are resistant to many chemicals and autoclavable
Varispenser® 2 and 2x bottle-top dispensers from Eppendorf fit on a variety of supply bottles and are chemical-resistant
Handle fuming acids with ease
Videos not loading, because cookies have been rejected. Change your
General handling
Videos not loading, because cookies have been rejected. Change your
Working with small vials
Videos not loading, because cookies have been rejected. Change your
How to Properly Handle Liquids from Bottles
Videos not loading, because cookies have been rejected. Change your
Product Information
- 1-channel, bottle top dispenser with recirculation valve and telescopic intake tube (length 170 – 330 mm)
- 2.5 – 25 mL
- incl. adapters GL 32, GL 38, S 40
You will find additional download material at the bottom of this page
Services (11)
Accessories (17)
Spare Parts (6)
11 Services
|
|
|
|
|
|
|
Show more Products
Added to Your Cart
Technical Data
Features
Technical Data
Specifications | Varispenser® 2x - 4967000049 |
---|
Product type | Dispensers |
Pipetting type | Positive displacement system |
Number of channels | 1-channel |
Volume range | 2.5 – 25 mL |
Volume increment | 0.5 mL |
Operating mode | mechanical |
Random measurement error |
2.5 mL: ± 1%; ± 25 µL 12.5 mL: ± 0.2%; ± 25 µL 25 mL: ± 0.1%; ± 25 µL |
Systematic measurement error |
2.5 mL: ± 5%; ± 125 µL 12.5 mL: ± 1%; ± 125 µL 25 mL: ± 0.5%; ± 125 µL |
Autoclavable | yes |
Open system (third party consumables) | yes |
Thread | GL 45 |
Bottle thread adapter incl. | GL 32, GL 38, S 401) |
Delivery content | Dispenser, Operating manual, Certificate, Telescopic aspirating tube (170 mm - 330 mm), Thread adapter (x3: 32 mm, 38 mm, 40 mm), Universal wrench |
1)buttress thread