TS 1000 single-channel dispensing tool
Technical Data
Features
Specifications | TS 1000 |
---|
Number of channels | 1-channel |
Volume range | 40 – 1,000 µL |
Testing Volume |
40 µL 100 µL 500 µL 1000 µL |
Rel. systematic error |
±5.0 % ±2.0 % ±1.0 % ±0.7 % |
Abs. systematic error |
±2.0 µL ±2.0 µL ±5.0 µL ±7.0 µL |
Abs. random error |
±0.6 µL ±1.0 µL ±1.0 µL ±1.5 µL |
Rel. random error |
±1.5 % ±1.0 % ±0.2 % ±0.15 % |