- Effective contamination protection: 140 °C sterilization routine
- Uniform atmosphere: Dual-channel IR-sensor (no TC-sensor), temperature uniformity verified (ISO 12880:2007-05), temperature and CO2 recovery after door opening < 5 min
- Easy cleaning: Stainless-steel chamber without seams, fan, ducts, covers etc. and removable water tray for safe and easy weekly cleaning and refill according to good cell culture practice
- Save money: Up to 25 % more usable space by fanless design, 3/4 shelves (max. load per shelf: 10 kg/22 lbs) + 2 access ports included, no costly internal expendable parts like fan-associated HEPA-filters that require bi-annual replacement
- Enhanced vibration and turbulence protection: No fan inside
- Low gas consumption: <1 L CO2/24 h (5 % CO2, w/o door opening)
- You need more? If you need more options (e.g. O2 control, humidity monitoring, copper interior), 180 °C disinfection, or future flexibility – consider the CellXpert® C170i
CA | CAD
CA | CAD
-
- Benchtop Centrifuges
- Floor-Standing Centrifuges
- Refrigerated Centrifuges
- Microcentrifuges
- Multipurpose Centrifuges
- High-Speed Centrifuges
- Ultracentrifuges
- Concentrator
- High-Speed and Ultracentrifugation Consumables
- Accessories
- Tubes
- Plates
- Device Management Software
- Sample and Information Management
CellXpert® C170
Product Information
- non-segmented inner door, handle right side
- 100 – 127 V/50 – 60 Hz (US/JP/South America), 3 shelves included
You will find additional download material at the bottom of this page
Technical Data
Features
Technical Data
Specifications | CellXpert® C170 |
---|
Capacity | 170 L (6.0 ft³) |
Power supply | 100 – 127 V, 50 – 60 Hz |
Dimensions (W × D × H) | 71.8 × 71.5 × 90.0 cm / 28.3 × 28.1 × 35.4 in |
Product weight | 113 kg / 249 lb |
High temperature disinfection (HTD) | 140 °C |
Sealed inner glass doors | yes |
Segmented door options | 4 segments |
Temperature control increment | ±0.1 °C |
Temperature stability at 37 °C | ±0.1 °C |
Temperature range | Ambient +4 °C to 50 °C |
Temperature uniformity | ±0.3 °C |
CO2 range | 0.1 – 20 % |
CO2 control increment | ±0.1 % |
CO2 stability at 5 % CO2 | ±0.1 % |
CO2 uniformity | ±0.1 % |
Number of shelves (standard/max) | 3 (4 with 4-segmented door)/8 |
Max. load/shelf | 10 kg |
Max. load/rack | 40 kg |
Data export interface | Ethernet |
Benchtop (B), Under Bench (U), Floor (F) or Stackable (S) | B, U, F, S (x2) |
Copper package | – |
Access port | 2 |
On-board data logging | – |
Water/Humidity monitoring | – |
BMS relay | yes |
Temperature range | Ambient +4 °C to 50 °C |
Temperature accuracy | ±0.4 °C |