- Heating rate: up to max. 5 °C/s
- Cooling rate above RT: max. 2.3 °C/s
- Control accuracy: ±0.15 °C
- Innovative 2D-Gradient for advanced PCR optimization
- X50 with Touchscreen can control up to 9 other units (connected cyclers do not have a touch screen interface)
- flexlid® concept: automatic height adjustment of the lid allows you to use all types of consumables
- Small footprint
- Connect up to 50 units when using the CycleManager X50 software
- Transfer your programs with constant parameters from other cycler models
- Warranty: 2 Year warranty
Mastercycler® X50a
Product Information
- 100 – 240 V/50 – 60 Hz (GB)
- aluminum block, 96-well or 0.1/0.2 mL tubes
- with touchscreen interface
You will find additional download material at the bottom of this page
Technical Data
Features
Technical Data
Specifications | Mastercycler® X50a |
---|
Unit can control 9 other units in network | ■ |
Unit can only be controlled | – |
Thermoblock | Aluminum |
Sample capacity | 96-well plate or 0.1/0.2 mL tubes |
96-well plate | ■ |
384-well plate | – |
0.1/0.2 mL tubes | ■ |
Heating technology | Six peltier elements |
2D-Gradient block | over 12 columns and 8 rows |
Temperature control mode | Fast, Intermediate, Standard, Safe |
Temperature control range of the block | 4 – 99 °C |
Gradient range | 1 – 30 °C |
Gradient temperature control range | 30 – 99 °C |
Block homogeneity 20 – 72 °C | ≤±0.2 °C |
Block homogeneity 95 °C | ≤±0.3 °C |
Control accuracy | ±0.15 °C |
Product type | PCR |
Heating rate | max. 5 °C/s |
Cooling rate | max. 2.3 °C/s |
Lid descent | flexlid® technology with thermal sample protection |
High pressure lid | – |
Lid temperature range | 37 – 110 °C |
Interfaces | Ethernet, USB |
Power supply | 100 – 240 V, 50 – 60 Hz |
Max. power consumption | 850 W |
Total weight | 11.5 kg |
Dimensions (W × D × H) | 27.5 × 43 × 33 cm / 10.8 × 16.9 × 13 in |
Remote Device Monitoring and Notifications (VisioNize®) | VisioNize® touch enabled |
Touchscreen Interface | yes |
Eppendorf ThermoMixer® |
|
Heating rate | − |
Cooling rate above RT | − |