- High centrifugation speed of up to 58,700 x g with fixed-angle rotor R27A (22,000 rpm)
- Maximum capacity of 6 L with 4 x 1.5 L unique triangular bottles
- Temperature range: -20 °C to 40 °C
- All rotors maintain 4 °C at maximum rotor speed (except R27A)
- High-resolution 5.7 inch color LCD with 2 different language options (EN/JPN)
- User management and documentation function
- Rotor cover detection sensor prevents rotors from spinning without rotor covers
- Automatic rotor ID detection and self-locking rotor system
- Variable deceleration slope function to protect sensitive samples (at <7,000 rpm)
- Biosafety Option (micro filter modification)
BN | BND
BN | BND
-
- Benchtop Centrifuges
- Floor-Standing Centrifuges
- Refrigerated Centrifuges
- Microcentrifuges
- Multipurpose Centrifuges
- High-Speed Centrifuges
- Ultracentrifuges
- Concentrator
- Accessories
- High-Speed and Ultracentrifugation Consumables
- Tubes
- Plates
- Device Management Software
- Sample and Information Management
Centrifuge CR22N
Technical Data
Features
Technical Data
Specifications | CR22N |
---|
Max. RCF | 58700 × g |
Speed | 300 – 22000 rpm |
Max. capacity | 4 × 1,500 mL |
Number of programs | 30 programs, with RCF(g) / g x sec function |
Noise level | 64 dB(A) |
Interfaces | USB port (operating history can be exported via USB port) |
Power supply | 220 – 240 V, 50/60 Hz |
Max. power consumption | 3200 W |
Dimensions (W × D × H) | 700 × 760 × 915 mm |
Footprint (dimensions w/o front panel, WxD) | 700 x 760 mm |
Product weight | 260 kg |
Cooling | refrigerated |
Separation example | High centrifugation speed of up to 58,700 x g with fixed-angle rotor R27A (22,000 rpm) |
Acceleration/Deceleration mode |
Acceleration: 9 modes Deceleration: 9 modes with coasting |
Drive unit | Induction Motor (Inverter) |
Refrigerator | Rotary inverter driven compressor (Refrigerant : R410A) |
Keypad | Touch sensitive LCD panel |
Safety devices | Door interlock, overspeed detection, imbalance detection, rotor cover detection, over-temperature detection |
Power requirement | Single phase: AC200/208/220/230/240 V ±10%, 30 A (50/60 Hz) |
Variable Deceleration Slope | yes (beside DCL mode, moderate deceleration or free coast is selectable) |
User lockout function | yes |
Operating Log Management |
LAN port: factory option or on-site modification PC and optional "himac LogManager Ver. 5.0 for Windows® (Network Edition)" is required Supporting GMP/GLP |
Economy mode | yes |
Continuous flow rotor | no |
Vacuum system | no |