DASGIP® Parallel Bioreactor System
Product Information
- for cell culture, max. 50 sL/h gassing
- 4-fold system with DASGIP® Bioblock
You will find additional download material at the bottom of this page
Technical Data
Technical Data
Specifications | DASGIP® System for Cell Culture |
---|
Application | Cell culture |
Number of parallel units | Up to 16 |
Software | DASware control, other DASware optional |
User Interface | Process computer with monitor |
Typical power consumption (4-fold system without process computer) |
480 W (at 230 V)/ 432 W (at 115 V) (595 W/552 W without DASGIP Bioblock) |
Dimensions (W × D × H, 4-fold system without process computer) | 140 x 70 x 53 cm (55 x 28 x 21 in) |
Typical weight |
78 kg (60 kg without DASGIP Bioblock) |
Bacteria/Yeast/Fungi | yes |
Mammalian/Animal | yes |
Insect | yes |
Interchangeable Vessels | yes |
Validatable | no |
Stem cells | yes |
Bioreactors |
|
Vessels | Glass/single-use vessels |
Sterilization |
Autoclavable/ pre-sterilized |
Total volume | 1.5 L to 2.3 L |
Agitation |
|
Drive | Direct/magnetic overhead drive |
Speed ranges |
30 – 1,250 rpm (standard)/ 100 – 1,600 rpm (optional) |
Impellers | Pitched-blade |
Gassing |
|
Gas supply | TMFC; overlay and/or sparger |
Standard gas flow rates | 0.1 – 50 sL/h, 0.1 – 40 sL/h CO2 (DASGIP MX4/4, other modules available) |
Standard gas mixing | Air, N2, O2 and/or CO2 |
Feeding |
|
Feed lines per vessel | Up to 8 |
Standard feed rates (depending on tube diameter) |
0.3 – 9.5 mL/h to 13 – 420 mL/h (DASGIP MP8) 10 – 70 mL/h to 0.4 – 5 L/h (DASGIP MP4) |
Monitoring and Control |
|
Temperature control | Heating and cooling integrated in DASGIP Bioblock (additional cooling fingers optional) |
Standard temperature range (at 25° C room temp.) | 5 K above cooling agent temperature – 99°C (DASGIP Bioblock)1) |
pH control | − |
DO control | Cascade (O2 concentration, gas flow rate), and other set-ups |
ORP (redox) measurement | - |
Level/foam | Optional |
OD measurement |
Optional (DASGIP OD4) |
Exhaust condensation | Water-cooled or liquid-free (Peltier with DASGIP EGC4) |
Exhaust analysis |
Optional (DASGIP GA4) |
1)Application-specific operating conditions apply. |