- Mechanical air-cushion pipette with standard 3-button operation (one operating element for volume selection, one for pipetting, and one for tip ejection)
- Ideal for accurate, easy, and ergonomic manual pipetting of aqueous solutions
- Feel the difference in weight and pipetting forces: Designed according to the Eppendorf PhysioCare Concept® principles
- Benefit from one of the lowest operating and tip ejection forces on the pipette market, helping to reduce hand and arm strain
- Enjoy minimal tip attachment forces due to a spring-loaded tip cone (available for all pipettes up to 1 mL)
- Temporarily adjust your pipette in seconds for better accuracy when pipetting challenging liquids like ethanol or when pipetting at high altitudes. Return to factory adjustment without calibration
- Autoclave the entire pipette or only the lower part according to your needs and applications to ensure a proper decontamination
- Premium material ensures a high UV and chemical resistance; special single-channel lower parts for aggressive chemicals available
- ACT®-labeled single-channel pipettes and epT.I.P.S.® pipette tips help you make an informed choice when it comes to sustainability
Eppendorf Research® plus
Product Information
- 1-channel
- 0,1 – 2,5 µL, dark gray, variable volume
- incl. epT.I.P.S.® Box 2.0 with 96 pipette tips
- ACT-labeled: Independent third-party product sustainability validation by My Green Lab
You will find additional download material at the bottom of this page
Pipette Tips (18)
Services (11)
Accessories (6)
Spare Parts (6)
|
|
|
|
|
|
|
|
|
|
Added to Your Cart
Technical Data
Features
Technical Data
Specifications | Eppendorf Research® plus - 3123000012 |
---|
ACT label | yes |
Product type | Pipettes |
Pipetting type | Air-cushion system |
Number of channels | 1-channel |
Volume range | 0,1 – 2,5 µL |
Color code | dark gray |
Volume increment | 0.002 µL |
Operating mode | mechanical |
Spring-loaded tip cone | yes |
Random measurement error |
0.1 µL: ± 12%; ± 0.012 µL 0.25 µL: ± 6%; ± 0.015 µL 1.25 µL: ± 1.5%; ± 0.01875 µL 2.5 µL: ± 0.7%; ± 0.0175 µL |
Systematic measurement error |
0.1 µL: ± 48%; ± 0.048 µL 0.25 µL: ± 12%; ± 0.03 µL 1.25 µL: ± 2.5%; ± 0.031 µL 2.5 µL: ± 1.4%; ± 0.035 µL |
Autoclavable | yes |
Consumables | epT.I.P.S.® |
Open system (third party consumables) | yes |
Delivery content | Pipette, epT.I.P.S.® Box 2.0 pipette tips, Operating manual, Certificate, Adjustment tool (Allen key with blue handle), Adjustment seal for temporary adjustment (x5), Grease for pipettes, Locking ring (to prevent spring action) |