- Autoclavable and calibrated based on ISO 8655-6
AR | ARS
AR | ARS
TS 300 single-channel dispensing tool
Technical Data
Features
Technical Data
Specifications | TS 300 |
---|
Number of channels | 1-channel |
Volume range | 20 – 300 µL |
Testing Volume |
20 µL 30 µL 150 µL 300 µL |
Rel. systematic error |
±4.0 % ±3.0 % ±1.0 % ±0.6 % |
Abs. systematic error |
±0.8 µL ±0.9 µL ±1.5 µL ±1.8 µL |
Abs. random error |
±0.5 µL ±0.45 µL ±0.6 µL ±0.9 µL |
Rel. random error |
±2.5 % ±1.5 % ±0.4 % ±0.3 % |