MENU
US | USD
US | USD
-
- Benchtop Centrifuges
- Floor-Standing Centrifuges
- Refrigerated Centrifuges
- Microcentrifuges
- Multipurpose Centrifuges
- High-Speed Centrifuges
- Ultracentrifuges
- Concentrator
- High-Speed and Ultracentrifuge Consumables
- Centrifuge Tubes
- Centrifuge Plates
- Device Management Software
- Sample and Information Management
No results found
Search Suggestions

When to Use a Fixed-angle Rotor or a Swing-bucket Rotor?
Lab Academy
- Centrifugation
- Centrifuges & Rotors
- Essay
Generally speaking, there are two types of centrifuge rotors available: fixed-angle rotors and swing-bucket rotors. Both types of rotors have advantages and disadvantages. Learn more about when to use which rotor type in this article.
Most fixed-angle rotors have bores at a fixed angle of 45° and the most compact pellet is formed at that angle. Smaller rotor angles result in more diffuse pellet areas.
The advantage of fixed-angle rotors is that they don’t have moving parts like swing-bucket rotors. Fixed-angle rotors are therefore exposed to much lower metal stress than swing-bucket rotors. This results in higher possible g-forces and thus shorter centrifugation times. The disadvantage of these rotors is that they provide only limited vessel capacity.
You will need a fixed-angle rotor when
The buckets of swing-bucket rotors swing out horizontally at a 90° angle, resulting in the pellets being located directly in the middle of the tube bottom. The strength of these rotors is that they enable high vessel capacity and high vessel flexibility due to the different adapter options available for them.
Pellets located in the middle of the tube bottom and not on the side of the tube wall as is the case with fixed-angle rotors can be an advantage as well, e.g., when low concentrations are expected and only very low volumes of buffer should be used for resolving the pellet in order to keep the sample concentration as high as possible.
Gradient centrifugation definitely requires a swing-bucket rotor so horizontal layers can form that remain in this position at the end of centrifugation.
A disadvantage of swing-bucket rotors is the high metal stress the moving buckets cause. This results in lower maximum g-forces and possibly requires longer centrifugation times.
You will need a swing-bucket rotor when you
The advantage of fixed-angle rotors is that they don’t have moving parts like swing-bucket rotors. Fixed-angle rotors are therefore exposed to much lower metal stress than swing-bucket rotors. This results in higher possible g-forces and thus shorter centrifugation times. The disadvantage of these rotors is that they provide only limited vessel capacity.
You will need a fixed-angle rotor when
- high g-forces are applied,
- compact pellets are needed, or
- a rather low number of vessels is used.
The buckets of swing-bucket rotors swing out horizontally at a 90° angle, resulting in the pellets being located directly in the middle of the tube bottom. The strength of these rotors is that they enable high vessel capacity and high vessel flexibility due to the different adapter options available for them.
Pellets located in the middle of the tube bottom and not on the side of the tube wall as is the case with fixed-angle rotors can be an advantage as well, e.g., when low concentrations are expected and only very low volumes of buffer should be used for resolving the pellet in order to keep the sample concentration as high as possible.
Gradient centrifugation definitely requires a swing-bucket rotor so horizontal layers can form that remain in this position at the end of centrifugation.
A disadvantage of swing-bucket rotors is the high metal stress the moving buckets cause. This results in lower maximum g-forces and possibly requires longer centrifugation times.
You will need a swing-bucket rotor when you
- have a high sample throughput,
- need a high vessel flexibility (adapter flexibility),
- are centrifuging plates
- need to have the pellet in the middle of the tube bottom,
- or are carrying out gradient centrifugation.
Read more
Read less
