General Product Support
Do you have a specific question and would like to contact us directly? Please use the appropriate contact form.
Repairs
Certification & qualification
Preventive maintenance & calibration
Applications
FAQ by product groups
Pipettes
Centrifuges
Mastercycler
epMotion
ULT
CO2 Incubator
Shaker
Photometry
ThermoMixer
Cell Manipulation
Pipettes
How often should pipettes be calibrated?
As part of his test equipment monitoring process, the operator determines the frequency of the calibration as well as the quantity of the testing volume and the number of measurements for each testing volume. This may vary depending on our how often the pipette is used, by how many operators the device is used, how aggressive the liquids are that are to be pipetted and the maximum permissible errors defined by the operator. However, there should be at least 1x inspection per year. Also refer to ISO 8655.
What do I need to remember before shipping the pipette?
The decontamination declaration must be completed in full. The pipette also needs to be packaged correctly and safely for transport to prevent any damage in transit.
There may be special customs requirements for electronic pipettes with a rechargeable battery if these are shipped abroad.
There may be special customs requirements for electronic pipettes with a rechargeable battery if these are shipped abroad.
To which address do I send the pipettes for calibration?
You can find the address here .
How long does the calibration process take?
The duration of the calibration process depends on which services were selected. As a rule, it is 3 to 5 days, except for customer-specific requirements.
Would you like more information about our services for pipettes? Please click here .
Centrifuges
Why should I have my centrifuge serviced at regular intervals?
To ensure the optimum performance and precision of your device. This also enables the early detection of any defects which can then be corrected easily. Performing a large number of cycles increases the susceptibility to rotor damage. Any damage can be detected at an early stage during a qualified inspection, i.e. the rotor can be replaced before any serious damage occurs.
How often does my centrifuge need to be serviced?
This depends on how intensively it is used and varies depending on the load on the device. However, it is recommended to have the centrifuge serviced at least once a year.
How often does the OQ need to be carried out for a centrifuge?
The company is responsible for technical checks of devices.
Eppendorf recommends regular device qualifications depending on the intensity of use, however, they should be carried out at least once a year.
Eppendorf recommends regular device qualifications depending on the intensity of use, however, they should be carried out at least once a year.
How can I clean my centrifuges myself?
Use a neutral cleaning agent for cleaning. We recommend using alcohol or alcohol-based disinfectants for disinfection. Please ensure that the applicable legal regulations and guidelines are observed when selecting the disinfection method. You can find a detailed description in the video "Centrifuge Maintenance - The basis for your lab safety" or in the operating manual which can be downloaded from our website.
Would you like more information about our services for centrifuges? Please click here .
Mastercycler
Why should a temperature verification be carried out?
To ensure reproducible measured values and results we recommend thermal temperature verification.
What does temperature verification mean?
There are single-channel and multi-channel measuring systems. They provide the option to check whether the set block temperature is running according to the required specifications. If it is found that the values deviate from the specifications they can be adjusted.
How can my device be cleaned and which cleaning agents and disinfectants can be used for cleaning and disinfection?
Switch off the device, disconnect the mains/power plug and let it cool down. Make sure that no liquid can get into the inside of the housing. When cleaning has been completed, the device should only be connected when it is fully dry on the inside and on the outside.
Do not use any aggressive chemicals on the device or its accessories such as strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol. If the device is contaminated with aggressive chemicals, clean it immediately using a mild cleaning agent. Use a lint-free cloth and a mild, soapy laboratory cleaner.
Select disinfection methods that comply with the legal regulations and guidelines for your area of application. If you have any questions in this regard, please contact the local Eppendorf Service Team.
Clean your device at regular intervals.
Please refer to your operating manual for further information - Maintenance chapter.
Do not use any aggressive chemicals on the device or its accessories such as strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol. If the device is contaminated with aggressive chemicals, clean it immediately using a mild cleaning agent. Use a lint-free cloth and a mild, soapy laboratory cleaner.
Select disinfection methods that comply with the legal regulations and guidelines for your area of application. If you have any questions in this regard, please contact the local Eppendorf Service Team.
Clean your device at regular intervals.
Please refer to your operating manual for further information - Maintenance chapter.
Would you like more information about our services for mastercycler? Please click here .
epMotion
Why should I have my epMotion serviced at regular intervals?
To ensure the optimum performance and precision of your device. This also enables the early detection of any defects which can then be corrected easily. Only precise and accurate devices ensure successful liquid transfers!
How often do dispensing tools need to be calibrated?
As part of his test equipment monitoring process, the operator determines the frequency of the calibration as well as the quantity of the testing volume and the number of measurements for each testing volume. This may vary depending on how often the device is used, by how many operators it is used, the aggressivity of the liquids used and the maximum permissible errors defined by the operator. The dispensing tools of your device will display after 200,000 runs that an inspection is recommended. However, there should be at least 1 calibration per year.
Why can the dispending tool be calibrated without the user unit?
The epMotion dispensing tool is calibrated with the help of the epMotion Tool Holder. The electronic parameter to control the tools of the epMotion are always the same in all epMotion units. Therefore all epMotion dispensing tool can be calibrated on different units than the user unit. The only exceptions are the TS10 and TM10 dispensing tool. They can only be calibrated with units controlled by epBlue Version 40.5.
Would you like more information about our services for an epMotion? Please click here .
ULT
How often does my freezer need to be serviced?
This depends on how intensively it is used and varies depending on the load of the device. However, we recommend to have it serviced at least once a year.
What measures can I take to care for the device?
The air filter can be cleaned approx. every three months (depending on the ambient conditions). In addition, if there is any ice build-up in the device this should be removed. The seals should be cleaned once a month and, if required, the device can be defrosted and cleaned on the inside and on the outside. Please refer to your operating manual for further information - Maintenance chapter.
How do I defrost my ULT-freezer?
When it has been operating for longer periods, defrosting may be required. You can find further information on how to proceed in the operating manual of your Eppendorf freezer.
Would you like more information about our services for ULT? Please click here .
CO2 Incubator
How often does my incubator need to be serviced?
This depends on how intensively it is used and varies depending on the load on the device. However, we recommend to have it serviced at least once a year.
What do you need to look out for before carrying out a service?
The device needs to be decontaminated first. The device must be freely accessible and all samples removed from the device. You should also ensure that the device has cooled down already.
How often does the OQ need to be carried out for an incubator?
The company is responsible for technical checks of devices. Eppendorf recommends regular device qualifications depending on the intensity of use, however, they should be carried out at least once a year.
How can my incubator be cleaned/ disinfected?
Use a solution of 70% isopropanol (isopropyl alcohol) and 30% distilled water as a disinfectant. Follow the cleaning guidelines carefully that are described in your user manual.
Which parts of the incubator should be replaced at regular intervals?
Replace the HEPA filter of the CO2 supply line and the HEPA filter with automatic zero setting * at least once a year. (* Auto-zero filter is only used in R-models.)
Would you like more information about our services for CO2 incubators? Please click here .
Lab Shakers
Why should I have my Orbital Shaker serviced at regular intervals?
As with all products with moving parts, wear and tear is inevitable and can affect how consistently the product operates over a long period of time. Eppendorf products are engineered to minimize wear and tear, but a periodic service allows early identification and tracking of excessive wear before product performance is outside of acceptable tolerances. Regular servicing may also be a regulatory requirement of our users.
How often does my orbital shaker need to be serviced?
Eppendorf recommends a service once every 12 months. However, the regulatory requirements of our users vary greatly and can often be more frequent.
Would you like more information about our services for lab shakers? Please click here .
Photometry
What is checked by the qualification service and why is it important for a photometer?
In functional qualification (OQ), comparative measurements are carried out against certified reference materials. These filters are traceable to reference standards of the National Institute of Standards and Technology (NIST, Gaithersburg, MD, USA) and are used to check the systematic and random error of the photometer and its wavelengths 260 and 280 nm.
For the Eppendorf BioPhotometers, the number of flashes of the xenon lamp is defined and set using a special service filter.
A check of the light path is necessary in order to determine any impurities in the cuvette shaft.
Only through regular calibration can compliance with the manufacturer's specifications be checked and correcting adjustments be made to the instrument if necessary
For the Eppendorf BioPhotometers, the number of flashes of the xenon lamp is defined and set using a special service filter.
A check of the light path is necessary in order to determine any impurities in the cuvette shaft.
Only through regular calibration can compliance with the manufacturer's specifications be checked and correcting adjustments be made to the instrument if necessary
Would you like more information about our services for photometry? Please click here .
ThermoMixer
How often does my ThermoMixer need to be serviced?
This depends on how intensively it is used and varies depending on the load on the device. However, we recommend to have it serviced at least once a year.
What does the temperature verification include and how often does it need to be carried out?
To ensure that your ThermoMixer delivers reproducible and correct results, temperature verification and adjustment - if required - are strongly recommended. Temperature verification is a very accurate and precise method and complies with all laboratory requirements with regard to documentation and detection in accordance with national and international standards.
How can my device be cleaned and which cleaning agents and disinfectants can be used for cleaning and disinfection of ThermoMixer C, ThermoMixer F and ThermoStat C?
Switch off the device, disconnect the mains/power plug and make sure that no liquid can get into the inside of the housing. When cleaning has been completed, the device should only be connected when it is fully dry on the inside and on the outside.
Do not use any aggressive chemicals on the device or its accessories such as strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol. If the device is contaminated with aggressive chemicals, clean it immediately using a mild cleaning agent. Use a lint-free cloth and a mild, soapy laboratory cleaner and distilled water.
Clean the housing and the thermoblock regularly.
Please refer to your operating manual for further information - Maintenance chapter.
Do not use any aggressive chemicals on the device or its accessories such as strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol. If the device is contaminated with aggressive chemicals, clean it immediately using a mild cleaning agent. Use a lint-free cloth and a mild, soapy laboratory cleaner and distilled water.
Clean the housing and the thermoblock regularly.
Please refer to your operating manual for further information - Maintenance chapter.
Would you like more information about our services for ThermoMixers? Please click here .
Cell Manipulation
Why should I have my microinjector/ micromanipulator serviced at regular intervals?
To ensure the optimum performance and precision of your device. This also enables the early detection of any defects which can then be corrected easily. Only precise and accurate devices ensure a successful cell manipulation!
Which consumables should be replaced at regular intervals?
O-rings of FemtoJet 4 injection tube and grip head, consumption indicator depending on intensity of use.
How can I see that the O-rings need to be replaced?
If leaks become noticeable on the grip head, they need to be replaced.
How can my device be cleaned and which cleaning agents and disinfectants can be used for cleaning and disinfection?
Switch off the device, disconnect the mains/power plug and make sure that no liquid can get into the inside of the housing. When cleaning has been completed, the device should only be connected when it is fully dry on the inside and on the outside.
Do not use any aggressive chemicals on the device or its accessories such as strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol. If the device is contaminated with aggressive chemicals, clean it immediately using a mild cleaning agent. Use a lint-free cloth and a mild, soapy laboratory cleaner.
Select disinfection methods that comply with the legal regulations and guidelines for your area of application. If you have any questions in this regard, please contact the local Eppendorf Service Team.
Clean your device at least every 4 weeks.
Please refer to your operating manual for further information - Maintenance chapter.
Do not use any aggressive chemicals on the device or its accessories such as strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol. If the device is contaminated with aggressive chemicals, clean it immediately using a mild cleaning agent. Use a lint-free cloth and a mild, soapy laboratory cleaner.
Select disinfection methods that comply with the legal regulations and guidelines for your area of application. If you have any questions in this regard, please contact the local Eppendorf Service Team.
Clean your device at least every 4 weeks.
Please refer to your operating manual for further information - Maintenance chapter.
Would you like more information about our services for cell manipulation? Please click here .