- Check of installation environment
- Check of electrical/gas supply and fittings
- Replacement of in-line gas and auto-zero filter
- Check of door seals and hinges
- Check of chamber gas, temperature and humidity sensors, and alarm functionality
- Check of incubator shelf and racking system, and humidity trays
- Check of sensor covers
- Check of wiring system, PCB's, display and controller
- Validation and adjustment of operating parameters in acc. to specifications
AR | ARS
AR | ARS
CO2 Incubator Services - Preventive Maintenance
Features